OEM - CONVERTER – CNDX, Conductivity interface board calibration, information and more...

 

Conductivity interface board accommodating conductivity cell K=1 with 10K@25C thermistor interface board, with built-in temperature compensation and power regulator. Wiring, calibrating, information and more.

 

The OEM custom board is provided with wires (option) for remote applications. Simply connect it to your PLC or Controller and control the conductivity of your aqua solutions.

 

 

 

 

OEM - Conductivity interface board drawing

 

Drawing information

 

1. 9V power regulator (please be advised Hot!)

2. Power indicator

3. Zero Calibration

4. Calibration

5. &  6. Output voltage-devider, factory setting (10:1)

7. Cell and thermistor connector (MINI-DIN-6)

8. Stainless Steel (SS) tube encapsulating 10K-Ohms thermistor

 

Note: All on board trimmers (P3, P4 and P5) are multi-turn trimmers facilitates precise calibration. P5 is factory calibrate and should not recalibrate by the user.

 


 

Board mounting

 

Mounted easily via four or five 0.125" mounting holes. The board design to fit in EID-1599 series enclosure. Use 4-40 screws 0/25 long to mount it to the enclosure

 

Note: Wires, enclosure and five mounting screws are sold separately.

 

 

 

Board with conductivity cell and TC via Terminal Block (TB) mounting

 

Mounted easily via four or five 0.125" mounting holes. The board design to fit in EID-1599 series enclosure. Use 4-40 screws 0/25 long to mount it to the enclosure. Mount the enclosure via DIN-Rail mounting clip. (See application note 48 for more details.

 

Note: Wires, enclosure and five mounting screws are sold separately.

 

* For EID-1599 enclosure series click hear

 

 

 

Board with conductivity cell without TC via Terminal Block (TB) mounting

 

Note: 1%, 0.25W resistor connected across the TC input to insure 10K Ohms resistance at 25C. this option some times required when the sample is in water bath with constant temperature.

 

Mounted easily via four or five 0.125" mounting holes. The board design to fit in EID-1599 series enclosure. Use 4-40 screws 0/25 long to mount it to the enclosure. Mount the enclosure via DIN-Rail mounting clip. (See application note 48 for more details.

 

Note: Wires, enclosure and five mounting screws are sold separately.

 

* For EID-1599 enclosure series click hear

 

 

 


 

Board power supply information

 

Power via 11-24VDC or 11-16VAC or 12V car battery.  

 

Note: Power down transformer sold separately.

 

* For our 110V or 220V to 12VAC Power down transformer (Class 2 Type Transformers EID-BP series) click Here

 


 

Conductivity cell information

 

The board is designed to measure aqua solutions via conductivity cell with K=1 constant and 10K thermistor.

 

* For our big selection of conductivity cells click hear

 

Note: Conductivity cells should be calibrated using a standard solution before using.  When selecting a standard, choose one that has the approximate conductivity of the solution to be measured.  After using the probe for some time, you might consider cleaning the plates with a mild liquid soap or acetone - Note:  Do not use abrasive materials in cleaning.

 


 

Calibrations

 

1. Connect conductivity cell K=1 with Thermal Compensation TC (thermistor of 10K ohms @ 25 degrees centigrade) to the board (see below drawings).

 

Note: Any non–used wire, as well as shielded wire, can be connected to the Non-Connected (NC) pins of the MINI-DIN-6 connector.

 

 

 

Please be advised that wire connections to the board can be done either via the MINI-DIN-6 connector (PS/2 like connector), or wires can be soldered directly to the board (corresponding connector pins).

 

2. Connect and power the board via 11-24VDC or 11-16VAC or 12V car battery.

 

3. Short switch wires. Ideally it’s done via command from the PLC through dedicated relay. Switch will be sufficient, however, all that we recommend is that you power the board for about 30 – 90 seconds. At the same time, take a reading and turn the board off. This will extend the conductivity cell life (reduce build-up).

 

4. While the cell is dry (not with contact with any liquid) in room temperature (25C), set P3 trimmer (10K ohms trimmer) so the output voltage is 0.000V. This function is also known as zero calibration.

 

5. Insert the cell into buffer of 1413uS (NIST traceable) at 25C, and set the output voltage via P4 (10K ohms trimmer) so that the output voltage is 141.3mV. This function is also known as calibration.

 

6. Clean and dry the cell.

 

7. Insert the cell into unknown aqua-solution and get corresponding voltage readings.

 

Note: V-Out is scaled down in a ratio of 10:1 via R31 and P5 i.e. 1.413V to 141.3mV if you don’t need this function, shorten P5 (100K trimmer) pins.

 


 

Features

 

  • Built-in power LED indicator.

  • Zero output multi-turn trimmer adjustment.

  • Calibration multi-turn trimmer adjustment.

  • On board multi-turn trimmers facilitates precise calibration of output voltage

  • Wiring for remote power switch.

  • Power via 12 or 24V DC or 12 car battery (HVAC industrial standard).

  • Converts uS/cm or Voltage.

  • Built in temperature compensation for aqua solutions.

  • Accommodate any EID conductivity cell K=1 with 10K thermistor.

  • Can be customized to accommodate your requirement.

  • Industrial Standard output, ideal for remote sensing applications, such as water cooling towers.

  • Linear voltage output.

  • Built in current alternating cell driver oscillator providing long lasting cell.


 

Specifications

  • Input                                      Rated to 0-4000 uS/cm

  • Output Voltage                         0 – 4VDC standard, calibrated for 1413mV with 1413 uS/cm input @ 25C

  • PCB material                             FR4 (UL approved)

  • Input Filter                               L.P.F. 10Hz

  • Max power consumption              0.25W

  • Temp. Compensation:                set for 10K ohms thermistor

  • Storage temperature                  40 - 70   [F]

  • Operating temperature                32 - 140 [F] (0 to 60°C) @ 95% RH

  • Low board DC voltage                 9 [V]

  • Power indicator                          LED Red

  • Adjustment                               Via trimmers

  • Mounting                                  Four 0.125” holes

  • Manufactured                            100% in USA


Applications:

  • Water towers

  • Pool TDS

  • Boiler blow-down

  • Reverse osmosis

  • Rinse baths

  • Waste water treatment

  • Desalination

  • Level Detection

  • Salinity testing

  • TDS testing

  • Water treatment


  •     •